This calibration procedure applies to all models (34460A/61A/65A/70A).
Review Test Considerations and Gain Calibration Considerations before beginning this procedure.
Configuration: 4-Wire Ohms
This procedure adjusts the gain for the 2-wire and 4-wire Ohms functions, and the offset compensated Ohms function.
![]() |
Each range takes less than 4 seconds to complete. |
Instrument Settings | ||
---|---|---|
Input |
Function | Range |
100 Ω |
4-wire Ohms |
100 Ω |
1 kΩ | 1 kΩ | |
10 kΩ | 10 kΩ | |
100 kΩ | 100 kΩ | |
1 MΩ | 1 MΩ | |
10 MΩ | 10 MΩ | |
100 MΩ | 2-wire Ohms | 100 MΩ |
Next Procedure |
![]() |